# ISO 13485:2016 Requirements Breakdown
|
|
|
|
This document provides a comprehensive breakdown of ISO 13485:2016 requirements for medical device Quality Management Systems (QMS).
|
|
|
|
## Table of Contents
|
|
|
|
1. [Clause 4: Quality Management System](#clause-4-quality-management-system)
|
|
2. [Clause 5: Management Responsibility](#clause-5-management-responsibility)
|
|
3. [Clause 6: Resource Management](#clause-6-resource-management)
|
|
4. [Clause 7: Product Realization](#clause-7-product-realization)
|
|
5. [Clause 8: Measurement, Analysis and Improvement](#clause-8-measurement-analysis-and-improvement)
|
|
|
|
## Clause 4: Quality Management System
|
|
|
|
### 4.1 General Requirements
|
|
|
|
#### 4.1.1 QMS Requirements
|
|
- Establish, document, implement, and maintain a QMS
|
|
- Maintain its effectiveness in accordance with ISO 13485
|
|
- Document the QMS processes and their interactions
|
|
|
|
#### 4.1.2 Process Approach
|
|
- Identify processes needed for the QMS
|
|
- Determine sequence and interaction of these processes
|
|
- Determine criteria and methods for effective operation and control
|
|
- Ensure availability of resources and information
|
|
- Monitor, measure, and analyze processes
|
|
- Implement actions to achieve planned results and maintain effectiveness
|
|
|
|
#### 4.1.3 Outsourced Processes
|
|
- Control any QMS process that is outsourced
|
|
- Ensure control is documented in the QMS
|
|
- Outsourcing does not relieve the organization of responsibility
|
|
|
|
#### 4.1.4 General QMS Requirements
|
|
- Establish, document, implement, and maintain QMS requirements per ISO 13485
|
|
- Include requirements for medical devices and applicable regulatory requirements
|
|
- Establish documented procedures for QMS activities
|
|
|
|
#### 4.1.5 Risk Management
|
|
- Establish documented requirements for risk management in product realization
|
|
- Maintain risk management records
|
|
- Ensure risk management is conducted according to documented requirements
|
|
|
|
#### 4.1.6 Software Validation
|
|
- Validate computer software applications used in QMS
|
|
- Validation must be conducted prior to initial use and after changes
|
|
- Establish documented approach including:
|
|
- Risk associated with the software application
|
|
- Validation activities
|
|
- Acceptance criteria
|
|
- User responsibilities
|
|
- Validation records
|
|
|
|
### 4.2 Documentation Requirements
|
|
|
|
#### 4.2.1 General Documentation
|
|
QMS documentation must include:
|
|
- Quality policy and quality objectives
|
|
- Quality manual
|
|
- Documented procedures and records required by ISO 13485
|
|
- Documents required by organization for effective processes
|
|
- Records required by ISO 13485
|
|
- Medical device files as required by applicable regulatory requirements
|
|
|
|
#### 4.2.2 Quality Manual
|
|
Establish and maintain a quality manual that includes:
|
|
- Scope of the QMS with details and justification for exclusions
|
|
- Documented procedures or reference to them
|
|
- Description of interaction between QMS processes
|
|
- Structure of documentation used in the QMS
|
|
|
|
#### 4.2.3 Medical Device File
|
|
Establish and maintain a medical device file for each type or family that includes:
|
|
- General description, intended use/purpose
|
|
- Label and instructions for use specifications
|
|
- Specifications for product and/or manufacturing
|
|
- Specifications for procedures for purchasing, manufacturing, servicing
|
|
- Procedures for measuring and monitoring
|
|
- Installation requirements (if applicable)
|
|
- Risk management file(s)
|
|
- Verification and validation information
|
|
- Design and development file(s) when applicable
|
|
|
|
#### 4.2.4 Control of Documents
|
|
Establish documented procedure to:
|
|
- Approve documents before issue
|
|
- Review, update, and re-approve documents
|
|
- Ensure changes and current revision status are identified
|
|
- Ensure relevant versions are available at points of use
|
|
- Ensure documents remain legible and readily identifiable
|
|
- Control distribution of documents
|
|
- Prevent unintended use of obsolete documents
|
|
- Apply suitable identification if retained for any purpose
|
|
|
|
Document changes must:
|
|
- Be reviewed and approved by original function unless otherwise designated
|
|
- Have access to pertinent background information
|
|
- Be identified in the document or appropriate attachments
|
|
|
|
#### 4.2.5 Control of Records
|
|
Establish documented procedure for:
|
|
- Identification, storage, security, integrity, retrieval, retention time, and disposition
|
|
- Records must remain legible, readily identifiable, and retrievable
|
|
- Changes to records must remain identifiable
|
|
- Retention time must be at least the lifetime of the medical device
|
|
- Records may be stored on any media but must remain retrievable
|
|
|
|
## Clause 5: Management Responsibility
|
|
|
|
### 5.1 Management Commitment
|
|
Top management must provide evidence of commitment by:
|
|
- Communicating importance of meeting regulatory and customer requirements
|
|
- Establishing quality policy
|
|
- Establishing quality objectives
|
|
- Conducting management reviews
|
|
- Ensuring availability of resources
|
|
|
|
### 5.2 Customer Focus
|
|
- Determine customer requirements and regulatory requirements
|
|
- Ensure customer requirements are met to enhance satisfaction
|
|
- Maintain documented requirements related to the medical device
|
|
|
|
### 5.3 Quality Policy
|
|
- Appropriate to the organization
|
|
- Includes commitment to meet requirements and maintain QMS effectiveness
|
|
- Provides framework for quality objectives
|
|
- Communicated and understood within organization
|
|
- Reviewed for continuing suitability
|
|
|
|
### 5.4 Planning
|
|
|
|
#### 5.4.1 Quality Objectives
|
|
- Establish quality objectives at relevant functions and levels
|
|
- Must be measurable and consistent with quality policy
|
|
- Objectives must support conformity to product requirements
|
|
|
|
#### 5.4.2 QMS Planning
|
|
- Plan to meet general requirements and quality objectives
|
|
- Maintain QMS integrity when changes are planned and implemented
|
|
- Document planning
|
|
|
|
### 5.5 Responsibility, Authority and Communication
|
|
|
|
#### 5.5.1 Responsibility and Authority
|
|
- Define and communicate responsibilities and authorities
|
|
- Document roles that manage, perform, verify QMS work
|
|
- Identify interrelation of all personnel
|
|
|
|
#### 5.5.2 Management Representative
|
|
Appoint a member of management who:
|
|
- Ensures QMS processes are established, implemented, and maintained
|
|
- Reports to top management on QMS performance and improvement needs
|
|
- Ensures promotion of awareness of regulatory and customer requirements
|
|
|
|
#### 5.5.3 Internal Communication
|
|
- Ensure communication processes are established
|
|
- Ensure communication occurs regarding QMS effectiveness
|
|
|
|
### 5.6 Management Review
|
|
|
|
#### 5.6.1 General
|
|
- Review QMS at planned intervals (at least annually)
|
|
- Review to ensure continuing suitability, adequacy, and effectiveness
|
|
- Include assessment of opportunities for improvement
|
|
- Maintain records of management reviews
|
|
|
|
#### 5.6.2 Review Input
|
|
Include:
|
|
- Results of audits
|
|
- Customer feedback
|
|
- Process performance and product conformity
|
|
- Status of preventive and corrective actions
|
|
- Follow-up actions from previous reviews
|
|
- Changes affecting QMS
|
|
- Recommendations for improvement
|
|
- Applicable new or revised regulatory requirements
|
|
|
|
#### 5.6.3 Review Output
|
|
Include decisions and actions related to:
|
|
- Improvements to QMS effectiveness and processes
|
|
- Product improvements related to customer requirements
|
|
- Resource needs
|
|
- Changes necessary to maintain QMS effectiveness
|
|
|
|
## Clause 6: Resource Management
|
|
|
|
### 6.1 Provision of Resources
|
|
Determine and provide resources needed to:
|
|
- Implement and maintain QMS and its effectiveness
|
|
- Meet regulatory and customer requirements
|
|
|
|
### 6.2 Human Resources
|
|
|
|
#### 6.2 General
|
|
Personnel performing work affecting product quality must be competent based on:
|
|
- Education, training, skills, and experience
|
|
- Documented evidence of competence
|
|
|
|
#### 6.3 Infrastructure
|
|
Determine, provide, and maintain infrastructure including:
|
|
- Buildings, workspace, and associated utilities
|
|
- Process equipment (hardware and software)
|
|
- Supporting services
|
|
|
|
Infrastructure maintenance requirements:
|
|
- Document requirements including maintenance activities
|
|
- Document requirements when maintenance can affect product quality
|
|
- Maintain records of maintenance activities
|
|
|
|
### 6.4 Work Environment and Contamination Control
|
|
|
|
#### 6.4.1 Work Environment
|
|
- Determine and manage work environment needed for product conformity
|
|
- Document requirements for work environment
|
|
- Document requirements if work environment can adversely affect product quality
|
|
|
|
#### 6.4.2 Contamination Control
|
|
- When applicable to medical device, document requirements for control of contaminated or potentially contaminated product
|
|
- Establish special arrangements for control of contaminated product
|
|
|
|
## Clause 7: Product Realization
|
|
|
|
### 7.1 Planning of Product Realization
|
|
Plan and develop processes needed for product realization including:
|
|
- Quality objectives and requirements for the product
|
|
- Need to establish processes, documentation, and resources
|
|
- Required verification, validation, monitoring, measurement, inspection, handling, storage, distribution, and traceability
|
|
- Records to provide evidence of conformity
|
|
|
|
Risk management requirements:
|
|
- Establish documented requirements for risk management throughout product realization
|
|
- Maintain risk management records
|
|
|
|
### 7.2 Customer-Related Processes
|
|
|
|
#### 7.2.1 Determination of Requirements
|
|
Determine:
|
|
- Requirements specified by customer including delivery and post-delivery
|
|
- Requirements not stated but necessary for specified or intended use
|
|
- Applicable regulatory requirements
|
|
- Any additional requirements determined by organization
|
|
|
|
#### 7.2.2 Review of Requirements
|
|
- Review product requirements before commitment
|
|
- Ensure requirements are defined and documented
|
|
- Ensure differences are resolved
|
|
- Ensure ability to meet requirements
|
|
- Maintain records of review results and follow-up actions
|
|
|
|
#### 7.2.3 Communication
|
|
Establish and document effective arrangements for communication with customers concerning:
|
|
- Product information
|
|
- Inquiry, contract or order handling, amendments
|
|
- Customer feedback including complaints
|
|
- Advisory notices
|
|
|
|
### 7.3 Design and Development
|
|
|
|
#### 7.3.1 General
|
|
- Establish, document, and maintain design and development procedures
|
|
- Document design and development plan for each medical device
|
|
- Maintain design and development files
|
|
|
|
#### 7.3.2 Design and Development Planning
|
|
Plan and control design and development including:
|
|
- Stages of design and development
|
|
- Required review, verification, and validation activities
|
|
- Responsibilities and authorities
|
|
- Resources and interfaces
|
|
- Update plans as design progresses
|
|
- Document plans
|
|
|
|
#### 7.3.3 Design and Development Inputs
|
|
- Determine inputs relating to product requirements
|
|
- Include functional, performance, usability, and safety requirements
|
|
- Include applicable regulatory requirements and standards
|
|
- Include applicable outputs of risk management
|
|
- Include appropriate information from previous similar designs
|
|
- Review inputs for adequacy and completeness
|
|
- Resolve incomplete, ambiguous, or conflicting requirements
|
|
- Maintain records
|
|
|
|
#### 7.3.4 Design and Development Outputs
|
|
Provide outputs that:
|
|
- Meet design input requirements
|
|
- Provide appropriate information for purchasing, production, and service
|
|
- Contain or reference product acceptance criteria
|
|
- Specify characteristics essential for safe and proper use
|
|
- Document outputs and maintain as records
|
|
|
|
#### 7.3.5 Design and Development Review
|
|
- Conduct systematic reviews at suitable stages
|
|
- Evaluate ability to meet requirements
|
|
- Identify problems and propose actions
|
|
- Include representatives of functions concerned
|
|
- Maintain records including results and follow-up actions
|
|
|
|
#### 7.3.6 Design and Development Verification
|
|
- Perform verification per planned arrangements
|
|
- Ensure outputs meet input requirements
|
|
- Maintain records of verification results and follow-up actions
|
|
|
|
#### 7.3.7 Design and Development Validation
|
|
- Perform validation per planned arrangements
|
|
- Ensure product meets specified application or intended use
|
|
- Conduct validation before delivery or implementation
|
|
- Include validation under defined operating conditions
|
|
- Maintain records of validation results and follow-up actions
|
|
|
|
#### 7.3.8 Design and Development Transfer
|
|
- Document procedures for transfer to manufacturing
|
|
- Verify manufacturing output meets design output
|
|
- Ensure specification for materials, production, QC, servicing are appropriate
|
|
- Maintain records
|
|
|
|
#### 7.3.9 Control of Design and Development Changes
|
|
- Identify, document, and control changes
|
|
- Review, verify, validate, and approve changes before implementation
|
|
- Evaluate effects on constituent parts, in-process product, and delivered product
|
|
- Maintain records of changes, review results, and follow-up actions
|
|
|
|
#### 7.3.10 Design and Development Files
|
|
Establish and maintain design and development files for each type or family including:
|
|
- Design and development plan
|
|
- Design inputs
|
|
- Design outputs
|
|
- Design review, verification, validation records
|
|
- Design change records
|
|
- Risk management file
|
|
|
|
### 7.4 Purchasing
|
|
|
|
#### 7.4.1 Purchasing Process
|
|
- Ensure purchased product conforms to purchase information
|
|
- Establish documented processes for purchasing activities
|
|
- Establish criteria for evaluation and selection of suppliers
|
|
- Base criteria on ability to supply per organization's requirements
|
|
- Monitor supplier performance
|
|
- Maintain records of evaluations and follow-up actions
|
|
- Establish process for notifying suppliers of changed product requirements
|
|
|
|
#### 7.4.2 Purchasing Information
|
|
Purchasing information must include:
|
|
- Requirements for approval of product, procedures, processes, equipment
|
|
- Requirements for qualification of personnel
|
|
- Quality management system requirements
|
|
- Requirements for notification to organization of nonconforming product
|
|
- Agreement that suppliers provide notification of changes to purchased product
|
|
- Agreement that purchase information be communicated to sub-tier suppliers
|
|
|
|
#### 7.4.3 Verification of Purchased Product
|
|
- Establish and implement inspection or other activities to ensure conformity
|
|
- Document extent of verification
|
|
- Verify at supplier's premises when customer intends to perform verification at supplier
|
|
- Document verification arrangements and method of product release
|
|
|
|
### 7.5 Production and Service Provision
|
|
|
|
#### 7.5.1 Control of Production and Service Provision
|
|
Plan and carry out production under controlled conditions including:
|
|
- Availability of documented procedures and work instructions
|
|
- Availability of suitable infrastructure and work environment
|
|
- Availability of monitoring and measuring equipment
|
|
- Availability and use of suitable monitoring and measuring activities
|
|
- Implementation of product release, delivery, and post-delivery activities
|
|
- Implementation of defined operations for labelling and packaging
|
|
- Procedures for servicing if applicable
|
|
|
|
Document requirements for:
|
|
- Control of product cleanliness if applicable
|
|
- Control during installation and verification if applicable
|
|
|
|
#### 7.5.2 Cleanliness of Product
|
|
Document requirements if:
|
|
- Product is cleaned per specified requirements before sterilization and/or use
|
|
- Product cannot be cleaned before sterilization
|
|
- Product is supplied non-sterile to be cleaned and then sterilized
|
|
|
|
Establish requirements for product hygiene in manufacturing, handling, and storage.
|
|
|
|
#### 7.5.3 Installation Activities
|
|
If applicable:
|
|
- Document requirements for installation and verification
|
|
- Maintain records of installation and verification
|
|
|
|
#### 7.5.4 Servicing Activities
|
|
If servicing is specified requirement:
|
|
- Establish documented procedures, reference materials, and measurements for servicing
|
|
- Analyze records of servicing for feedback into post-production phase
|
|
- Maintain records of servicing activities
|
|
|
|
#### 7.5.5 Particular Requirements for Sterile Medical Devices
|
|
Maintain records of process parameters for sterilization of each batch.
|
|
|
|
#### 7.5.6 Validation of Processes
|
|
Validate processes where resulting output cannot be verified by subsequent monitoring or measurement, including:
|
|
- Defined criteria for review and approval
|
|
- Approval of equipment and qualification of personnel
|
|
- Use of specific methods, procedures, and acceptance criteria
|
|
- Requirements for records
|
|
- Revalidation including criteria for revalidation
|
|
- Approval of changes to process
|
|
|
|
Document requirements for validation of:
|
|
- Computer software used in production and service provision
|
|
- Sterilization processes
|
|
- Aseptic processing
|
|
- Clean room requirements if applicable
|
|
|
|
#### 7.5.7 Particular Requirements for Validation of Processes for Sterilization and Sterile Barrier Systems
|
|
Maintain records of validation of:
|
|
- Sterilization processes for each batch
|
|
- Sterile barrier systems
|
|
|
|
#### 7.5.8 Identification
|
|
- Establish documented procedures for product identification throughout realization
|
|
- Identify product by suitable means
|
|
- Maintain records of identification where traceability is a requirement
|
|
|
|
#### 7.5.9 Traceability
|
|
|
|
##### 7.5.9.1 General
|
|
Establish documented procedures defining extent of traceability including:
|
|
- Distribution and location of medical device
|
|
|
|
##### 7.5.9.2 Particular Requirements
|
|
Document procedures to maintain records of:
|
|
- Name and address of shipping package consignee
|
|
- Identification of quantity shipped
|
|
- Include requirements of applicable regulatory requirements
|
|
- Maintain traceability records for defined period
|
|
|
|
#### 7.5.10 Customer Property
|
|
- Exercise care with customer property while under organization's control
|
|
- Identify, verify, protect, and safeguard customer property
|
|
- Record and report to customer if lost, damaged, or unsuitable
|
|
- Maintain records
|
|
|
|
#### 7.5.11 Preservation of Product
|
|
- Preserve product during internal processing and delivery
|
|
- Include identification, handling, packaging, storage, and protection
|
|
- Apply to constituent parts of product
|
|
- Document requirements for special handling if applicable
|
|
|
|
### 7.6 Control of Monitoring and Measuring Equipment
|
|
- Determine monitoring and measurement to be undertaken
|
|
- Determine monitoring and measuring equipment needed
|
|
- Establish documented procedures for:
|
|
- Calibration or verification at specified intervals before use
|
|
- Adjustment or re-adjustment as necessary
|
|
- Identification to enable determination of calibration status
|
|
- Safeguarding from adjustments that would invalidate calibration
|
|
- Protection from damage and deterioration
|
|
- Assess and record validity of previous results when found not to conform
|
|
- Maintain records of calibration and verification
|
|
- Confirm ability of computer software to satisfy intended application when used
|
|
- Undertake confirmation before initial use and reconfirm as necessary
|
|
|
|
## Clause 8: Measurement, Analysis and Improvement
|
|
|
|
### 8.1 General
|
|
- Plan and implement monitoring, measurement, analysis, and improvement processes
|
|
- Demonstrate product conformity
|
|
- Ensure QMS conformity
|
|
- Maintain QMS effectiveness
|
|
- Include determination of applicable methods including statistical techniques
|
|
|
|
### 8.2 Monitoring and Measurement
|
|
|
|
#### 8.2.1 Feedback
|
|
Establish documented procedure for feedback including early warning system for:
|
|
- Post-production information including complaints
|
|
- Requirements for reporting to regulatory authorities
|
|
- Use as potential input to risk management for monitoring and maintaining product requirements
|
|
- Use as potential input for corrective and preventive action
|
|
|
|
#### 8.2.2 Complaint Handling
|
|
Establish documented procedures for timely complaint handling including:
|
|
- Requirements and responsibilities for receiving, recording, and evaluating complaints
|
|
- Requirements and responsibilities for handling, investigating, and evaluating complaints
|
|
- Requirements and responsibilities for reporting complaint information to regulatory authorities
|
|
- Requirements for informing customer of organization's actions
|
|
- Requirements to ensure complaint information not handled by organization is transferred to organization
|
|
- Maintain records of complaints and investigations
|
|
|
|
#### 8.2.3 Reporting to Regulatory Authorities
|
|
Establish documented procedures to:
|
|
- Provide notification to regulatory authorities per applicable requirements
|
|
- Provide advisory notices per applicable requirements
|
|
- Maintain records of reporting
|
|
|
|
#### 8.2.4 Internal Audit
|
|
- Conduct internal audits at planned intervals
|
|
- Determine if QMS conforms to ISO 13485 and organization's requirements
|
|
- Determine if QMS is effectively implemented and maintained
|
|
- Plan audit program considering importance of processes, changes, and previous results
|
|
- Define audit criteria, scope, frequency, and methods
|
|
- Establish documented procedure for audits including responsibilities, requirements, and reporting
|
|
- Select objective and impartial auditors
|
|
- Maintain records of audits and results
|
|
- Identify need for corrections or corrective actions
|
|
- Conduct follow-up activities to verify implementation and effectiveness
|
|
|
|
#### 8.2.5 Monitoring and Measurement of Processes
|
|
- Apply suitable methods for monitoring and measurement of QMS processes
|
|
- Demonstrate ability to achieve planned results
|
|
- Implement corrections and corrective actions when planned results not achieved
|
|
- Maintain records
|
|
|
|
#### 8.2.6 Monitoring and Measurement of Product
|
|
- Monitor and measure product characteristics to verify conformity
|
|
- Conduct at appropriate stages per planned arrangements
|
|
- Maintain records showing conformity to acceptance criteria
|
|
- Record authority responsible for release
|
|
- Ensure product release and delivery not proceed until planned arrangements completed
|
|
- Allow release by relevant authority and customer when applicable
|
|
|
|
### 8.3 Control of Nonconforming Product
|
|
|
|
#### 8.3.1 General
|
|
- Ensure nonconforming product is identified and controlled
|
|
- Establish documented procedures for:
|
|
- Identification, documentation, evaluation, segregation, and disposition
|
|
- Notification to external parties
|
|
- Review of nonconforming product
|
|
- Maintain records of nonconformities and subsequent actions including concessions
|
|
|
|
#### 8.3.2 Actions in Response to Nonconforming Product Detected Before Delivery
|
|
Deal with nonconforming product by:
|
|
- Taking action to eliminate detected nonconformity
|
|
- Authorizing use, release, or acceptance under concession per applicable regulatory requirements
|
|
- Taking action to preclude original intended use or application
|
|
- Taking action appropriate to effects of nonconformity when detected after delivery or use
|
|
|
|
Maintain records of concessions and identify authority making concession.
|
|
|
|
#### 8.3.3 Actions in Response to Nonconforming Product Detected After Delivery
|
|
When nonconforming product detected after delivery or use:
|
|
- Take action appropriate to effects of nonconformity
|
|
- Establish documented procedure including notification requirements to regulatory authorities
|
|
- Maintain records
|
|
|
|
#### 8.3.4 Rework
|
|
Establish documented procedures for rework including:
|
|
- Requirements to evaluate potential effects on medical device
|
|
- Approval before implementation
|
|
- Records of results and actions including nonconformities and rework
|
|
- Re-verification after rework
|
|
- Documentation of rework procedure before rework begins
|
|
|
|
### 8.4 Analysis of Data
|
|
- Determine, collect, and analyze appropriate data from monitoring and measurement
|
|
- Evaluate where continual improvement of QMS effectiveness can be made
|
|
- Establish documented procedures for:
|
|
- Analysis of data to provide information on customer satisfaction
|
|
- Analysis of conformity to product requirements
|
|
- Analysis of characteristics and trends of processes and products including preventive action opportunities
|
|
- Analysis of suppliers
|
|
- Analysis of other relevant data including feedback and output from risk management
|
|
- Include use of statistical techniques if necessary
|
|
- Maintain records of analysis results
|
|
|
|
### 8.5 Improvement
|
|
|
|
#### 8.5.1 General
|
|
- Identify and implement changes to ensure and maintain QMS effectiveness
|
|
- Include use of quality policy, objectives, audit results, data analysis, corrective and preventive actions, and management review
|
|
|
|
#### 8.5.2 Corrective Action
|
|
- Establish documented procedures to:
|
|
- Review nonconformities including complaints
|
|
- Determine causes of nonconformities
|
|
- Evaluate need for actions to ensure nonconformities do not recur
|
|
- Plan and document actions needed and implement
|
|
- Document results of actions taken
|
|
- Review effectiveness of corrective actions taken
|
|
- Maintain records including investigation results and follow-up
|
|
|
|
#### 8.5.3 Preventive Action
|
|
- Establish documented procedures to:
|
|
- Determine potential nonconformities and their causes
|
|
- Evaluate need for action to prevent occurrence
|
|
- Plan and document actions needed and implement
|
|
- Document results of actions taken
|
|
- Review effectiveness of preventive actions taken
|
|
- Use appropriate sources of information including:
|
|
- Work processes and operations affecting product quality
|
|
- Concessions
|
|
- Analysis of data and risk management outputs
|
|
- Medical device performance data
|
|
- Records of nonconformities
|
|
- Maintain records including investigation results and follow-up
|
|
|
|
## Key Regulatory Updates
|
|
|
|
### FDA QMSR Harmonization (Effective February 2, 2026)
|
|
- FDA 21 CFR Part 820 has been harmonized with ISO 13485:2016
|
|
- Renamed to QMSR (Quality Management System Regulation)
|
|
- Medical Device File (MDF) replaces separate DHF, DMR, and DHR
|
|
- Organizations should prepare for transition to unified documentation approach
|
|
|
|
## References and Resources
|
|
|
|
This requirements breakdown is based on ISO 13485:2016, which was last reviewed and confirmed in 2025.
|
|
|
|
For additional guidance, refer to:
|
|
- ISO 13485:2016 standard document
|
|
- FDA Quality Management System Regulation (QMSR)
|
|
- Applicable regional regulatory requirements (EU MDR, Health Canada, etc.)
|